Heathkit Model IT-18
In-circuit / Out-of-circuit Transistor Tester
Calibration Procedure
· The meter pointer should be on the extreme left-hand mark on the meter scale as shown in Figure 1. If the pointer is not over this mark, adjust the zero-adjust screw at the bottom of the meter until the pointer is properly aligned.
· Install a fresh D cell (1-1/2 volt battery) in the battery holder. Make sure the positive (+) end of the battery is facing the positive (+) terminal of the battery holder (refer to Figure 2.)
· Make sure that any exposed metal parts of the test leads are not touching each other.
· Always recalibrate the meter when a new battery has been installed.
Ancillary Equipment Needed
180-ohm 5% resistor (brown-gray-brown-gold)
* * * * * * *
Initial Condition Adjustments
Refer to Figure 1.
( ) Set the POLARITY switch (NPN-OFF-PNP) to the OFF position.
( ) Set the BETA CAL. control maximum counter-clockwise.
( ) Set the BETA multiplier switch to the X1 position.
( ) Set the LEAKAGE control [on rear of FUNCTION switch] maximum counter-clockwise (when viewed from the rear of the panel).
* * * * * * *
( ) Connect a 180-ohm 5% (brown-gray-brown-gold) resistor between the collector (red) and the emitter (white) test leads.
· Do not allow the meter to go off scale to the right and remain this way for any length of time, as this could damage the meter movement.
( ) Set the POLARITY switch to the NPN position.
( ) Hold the FUNCTION switch (ICEO-BETA-ICBO) in the ICEO position.
( ) Adjust the LEAKAGE control (on the rear of the FUNCTION switch) until the meter indicates 5K (full scale) on the leakage scale of the meter.
( ) Set the FUNCTION switch to the BETA position.
( ) Set the POLARITY switch to the OFF position.
( ) Connect the 180-ohm resistor between the base (green) and collector (red) test leads.
( ) Set the POLARITY switch to the NPN position.
( ) Hold the FUNCTION switch in the ICBO position.
( ) The meter should read full scale without readjustment of the LEAKAGE control.
( ) Set the FUNCTION switch to the BETA position.
( ) Set the POLARITY switch to the OFF position.
( ) Disconnect the 180-ohm resistor from the test leads.
( ) Calibration procedure is complete.