Heathkit Model IP-32
Regulated Power Supply
Calibration Procedure
· With the power supply off, both meter pointers should be over the zero indication on each meter face. If they are not, turn the mechanical zero-adjust screws at the bottom of each meter until the pointers are properly aligned.
Ancillary Equipment Needed
VTVM (optional)
* * * * * * *
NOTES
· Remove all jumper wires between the chassis ground and common ground terminals.
· Remove all loads from the supply terminals.
· All chassis mounted control adjustments are made from the top, front to rear (see Pictorial 7).
* * * * * * *
ZERO SET & 400V SET ADJUSTMENTS
( ) Set the POWER switch (SW) to the OFF position.
( ) Set the C- OUTPUT control (CC) [front panel] maximum counter-clockwise.
( ) Set the B+ OUTPUT control (CD) [front panel] maximum counter-clockwise.
( ) Set the ZERO SET control (CA) [chassis, front right - first adjustment potentiometer] maximum clockwise.
( ) Set the 400 VOLT SET control (CB) [chassis, front right - second adjustment potentiometer] maximum counter-clockwise.
( ) Set the meter slide switch (SL) [front panel] to the B+ OUTPUT position.
( ) Set the POWER switch (SW) to the STANDBY position. Allow the tubes to come up to operating temperature.
( ) Set the POWER switch (SW) to the ON position.
· There should be 0 output indicated on the current meter.
Meter Calibration without a VTVM
( )* Set the high voltage B+ OUTPUT control (CD) maximum counter-clockwise.
( )* Adjust the ZERO SET control (CA) counter-clockwise until the voltage meter reads 0 V.
( )* Set the high voltage B+ OUTPUT control (CD) maximum clockwise.
( )* Adjust the 400 VOLT SET control (CB) clockwise until the voltage meter reads 400 V.
· Because the ZERO SET and the 400 VOLT SET controls interact with each other, it will be necessary to repeat the above four steps* until no meter changes are observed.
( ) Set the high voltage B+ OUTPUT control (CD) maximum counter-clockwise.
Meter Calibration using a VTVM
( ) Set the high voltage B+ OUTPUT control (CD) maximum counter-clockwise.
( ) Set the POWER switch (SW) to the STANDBY position.
( ) Set the VTVM on a voltage scale capable of reading 400 VDC.
( ) Connect the VTVM to the B+ output terminals.
( ) Set the POWER switch (SW) to the ON position.
* * * * * * *
( )* Set the high voltage B+ OUTPUT control (CD) maximum counter-clockwise.
( )* Adjust the ZERO SET control (CA) counter-clockwise until the panel voltage meter reads 0 V.
( )* Adjust the high voltage B+ OUTPUT control (CD) until the VTVM reads 400 VDC.
( )* Adjust the 400 VOLT SET control (CB) clockwise until the panel voltage meter reads 400 V.
· Because the ZERO SET and the 400 VOLT SET controls interact with each other, it will be necessary to repeat the above four steps* until no meter changes are observed.
( ) Set the B+ OUTPUT control (CD) maximum counter-clockwise.
( ) Set the POWER switch (SW) to the STANDBY position.
( ) Disconnect VTVM from the B+ output terminals.
( ) Set the POWER switch (SW) to the ON position.
NEGATIVE BIAS VOLTAGE VERIFICATION
( ) Set the C- OUTPUT control (CC) maximum counter-clockwise.
( ) Set the meter slide switch (SL) to the C- OUTPUT position, voltage meter should read 0 V.
( ) Set the C- OUTPUT control (CC) maximum clockwise.
( ) Voltage meter should read over -100 V.
( ) Set the C- OUTPUT control (CC) maximum counter-clockwise.
( ) Set the POWER switch to the OFF position.
( ) Set the meter slide switch (SL) to the B+ OUTPUT position.
( ) Calibration procedure is completed.