Heathkit Model IP-27
Low Voltage Power Supply
Calibration Procedure
· With the power supply off, the meter pointer should be over the zero indication on the meter face. If it is not, adjust the mechanical zero-adjust screw at the bottom of the meter until the pointer is properly aligned.
Ancillary Equipment Required
10 ohm 25 watts resistor (optional)
Differential voltmeter (optional)
* * * * * * *
NOTES
· Remove all jumper wires between the circuit grounds and chassis ground.
· Remove all loads from the supply terminals.
* * * * * * *
Front Panel Initial Conditions
( ) Set the COARSE VOLTAGE switch to the AC OFF position.
( ) Set the D.C. ON – RESET STANDBY switch to the RESET STANDBY position.
( ) Set the COARSE CURRENT switch to 1.5 A.
( ) Set the FINE VOLTAGE control to center of rotation.
( ) Set the FINE CURRENT control maximum clockwise.
( ) Set the CURRENT – VOLTAGE switch to the VOLTAGE position.
*Back Panel* Initial Conditions
( ) Set the *ZENER CURRENT* control maximum counter-clockwise.
( ) Set the *VOLTAGE CALIBRATE* control to center of rotation.
( ) Set the *DC REGULATION* control to 1/8 turns from full clockwise.
( ) Set the *METER* switch to the ZENER CURRENT position.
( ) Set the POWER switch to the ON position.
( ) Set the COARSE VOLTAGE switch to the .5 – 5 V position.
· When the *METER* switch is in the ZENER CURRENT position, the front panel meter will read a 5 ma (full scale) indication.
( ) Adjust the *ZENER CURRENT* control for a meter reading of approximately 4 ma.
( ) Set the *METER* switch to the NORMAL position.
( ) Set the DC ON / RESET-STANDBY switch to the DC ON position.
( ) Adjust the FINE VOLTAGE control and observe a meter swing from almost zero to full scale.
( ) Set the FINE VOLTAGE control maximum clockwise.
( ) Set the COARSE VOLTAGE switch to the 45 – 50 V position.
( ) Adjust the FINE VOLTAGE control for a full scale meter reading.
( )* Set the *METER* switch to the ZENER CURRENT position.
( )* Adjust the *ZENER CURRENT* control for a full scale meter reading.
· The ZENER CURRENT readings are not critical and may vary.
( )* Set the *METER* switch to the NORMAL position.
( )* Adjust the *VOLTAGE CALIBRATE* control for a full scale meter reading.
· Because the *ZENER CURRENT* and the *VOLTAGE CALIBRATE* controls interact with each other, it will be necessary to repeat the above four steps* until no meter changes are observed.
( ) Set the *METER* switch to the NORMAL position.
( ) Set the COARSE VOLTAGE switch to the AC OFF position.
DC REGULATION CONTROL ADJUSTMENT
(optional)
· If a differential voltmeter is not available, leave the DC REGULATION control as originally set and proceed to the FINAL ADJUSTMENTS section.
Front Panel Initial Conditions
( ) Set the CURRENT – VOLTAGE switch to the VOLTAGE position.
( ) Set the COARSE VOLTAGE switch to the 10 – 15 V position.
( ) Set the FINE VOLTAGE control maximum clockwise.
( ) Set the COARSE CURRENT switch to the 1.5a position.
( ) Set the FINE CURRENT control maximum clockwise.
( ) Set the D.C. ON – RESET STANDBY switch to the RESET STANDBY position.
( ) Connect a differential voltmeter to the output terminals
( ) Set the D.C. ON – RESET STANDBY switch to the D.C. ON position.
( )* Note the unloaded output voltage on the differential voltmeter.
( )* Set the D.C. ON – RESET STANDBY switch to the RESET STANDBY position.
( )* Connect a 10 ohm 25 watts resistor to the output terminals.
( )* Set the D.C. ON – RESET STANDBY switch to the D.C. ON position.
( )* Note the loaded output voltage on the differential voltmeter.
( )* Adjust the *DC REGULATION* control to match the meter readings.
( )* Set the D.C. ON – RESET STANDBY switch to the RESET STANDBY position.
( )* Disconnect the 10 ohm resistor from the output terminals.
· In order for the output voltage regulation to be the same both loaded and non-loaded, it will be necessary to repeat the above eight steps* until no meter changes are observed.
FINAL ADJUSTMENTS
( ) Set the FINE VOLTAGE control maximum counter-clockwise.
( ) Set the COARSE VOLTAGE switch to AC OFF.
( ) Set the D.C. ON – RESET STANDBY switch to the RESET STANDBY position.
( ) Set the POWER switch to the OFF position.
( ) Disconnect the resistive load.
( ) Calibration procedure is complete.