Heathkit Model IP-20
Transistorized Power Supply
Calibration Procedure
· With the power supply off, the meter pointer should be over the zero indication on the meter face. If it does not, adjust the mechanical zero-adjust screw at the bottom of the meter until the pointer is properly aligned.
Ancillary Equipment Needed
10 ohm 25 watts resistor (optional)
Differential Voltmeter (optional)
* * * * * * *
NOTES
· Remove all jumper wires between the circuit grounds and chassis ground.
· Remove all loads from the supply terminals.
· Always change the VOLTAGE RANGE switch slowly to avoid possible voltage transients that may damage the transistors.
* * * * * * *
Front Panel Initial Conditions
( ) Set the POWER switch [on back of unit] to the OFF position.
( ) Set the VOLTAGE RANGE switch to the AC OFF position.
( ) Set the CURRENT RANGE switch to the 1.5A position.
( ) Set the FINE VOLTAGE control to center of rotation.
( ) Set the CURRENT LIMIT control maximum clockwise.
( ) Set the METER switch to the VOLTAGE position.
( ) Remove the guard nut from the DC REGULATOR control.
( ) Set the DC REGULATION control to 1/8 turns from full counter-clockwise.
( ) Replace the guard nut on the DC REGULATOR control.
( ) Set the D.C. ON – RESET STANDBY switch to the RESET STANDBY position.
*Control Plate* Initial Conditions
( ) Set the *ZENER CURRENT* control maximum count-clockwise.
( ) Set the *VOLTAGE CALIBRATE* control to center of rotation.
( ) Set the *METER* switch to the ZENER CURRENT position.
( ) Set the POWER switch [on back of unit] to the ON position. Allow the tubes to come up to operating temperature.
( ) Set the VOLTAGE RANGE switch to the 5V position.
( ) Adjust the *VOLTAGE CALIBRATE* control for a 5 ma (full scale) meter reading.
( ) Set the *METER* switch to the NORMAL position.
( ) Set the D.C. ON – RESET STANDBY switch to the D.C. ON position.
· The 10 & 15 volt ranges use the 0 - 15 scale. The remaining ranges use the 0 - 50 scale.
( )* Adjust the FINE VOLTAGE control and observe the meter move from almost zero to full scale.
( )* Set the FINE VOLTAGE control maximum counter-clockwise.
· Repeat the above two steps* while slowly incrementing the VOLTAGE RANGE switch from the 5V to the 50V positions.
( ) Set the VOLTAGE RANGE switch to the 50V position.
( ) Set the FINE VOLTAGE control maximum clockwise.
( )* Adjust the *VOLTAGE CALIBRATE* control for a full scale meter reading.
( )* Set the *METER* switch to the ZENER CURRENT position.
( )* Adjust the *ZENER CURRENT* control for a full scale meter reading.
( )* Set the *METER* switch to the NORMAL position.
· Because the *VOLTAGE CONTROL* and the *ZENER CURRENT* control interact with each other, it will be necessary to repeat the above four steps* until no meter changes are observed.
( ) Set the *METER* switch to the NORMAL position.
( ) Decrement the VOLTAGE RANGE switch from the 50V to the AC OFF positions.
DC REGULATION CONTROL ADJUSTMENT
(optional)
· If a differential voltmeter is not available, leave the DC REGULATION control as originally set and proceed to the FINAL ADJUSTMENTS section.
Front Panel Initial Conditions
( ) Increment the VOLTAGE RANGE switch to the 15V position.
( ) Set the METER switch to the VOLTAGE position.
( ) Set the CURRENT RANGE switch to the 1.5A position.
( ) Set the CURRENT LIMIT control maximum clockwise.
( ) Remove the guard nut from the DC REGULATOR control.
( ) Set the D.C. ON – RESET STANDBY switch to the RESET STANDBY position.
( ) Connect a differential voltmeter across the output terminals.
( ) Set the D.C. ON – RESET STANDBY switch to the D.C. ON position.
( )* Measure the unloaded output voltage using the differential voltmeter.
( )* Set the D.C. ON – RESET STANDBY switch to the RESET STANDBY position.
( )* Connect a 10 ohm 25 watts resistor to the output terminals.
( )* Set the D.C. ON – RESET STANDBY switch to the D.C. ON position.
( )* Measure the loaded output voltage using the differential voltmeter.
( )* Adjust the *DC REGULATION* control to match the unloaded meter readings.
( )* Set the D.C. ON – RESET STANDBY switch to the RESET STANDBY position.
( )* Disconnect the 10 ohm resistor from the output terminals.
· In order for the output voltage regulation to be the same both loaded and non-loaded, it will be necessary to repeat the above eight steps* until no meter changes are observed.
( ) Disconnect the differential voltmeter from the output.
FINAL ADJUSTMENTS
( ) Replace the guard nut on the DC REGULATOR control.
( ) Set the D.C. ON – RESET STANDBY switch to the RESET STANDBY position.
( ) Set the FINE VOLTAGE control maximum counter-clockwise.
( ) Decrement the VOLTAGE RANGE switch slowly from 15V to AC OFF.
( ) Set the POWER switch [on back of unit] to the OFF position.
( ) Calibration procedure is complete.