Heathkit Model IB-1B
Impedance Bridge
Calibration Procedure
· The meter pointer should be on the center zero of the meter scale. If the pointer is not over this mark, adjust the zero-adjust screw at the bottom of the meter until the pointer is properly aligned.
Ancillary Equipment Needed
Frequency Counter
550-ohm 1% resistor
· Mount the 1% resistor on a dual banana plug to use as a calibration standard.
* * * * * * *
OSCILLATOR ADJUSTMENT
( ) Set the GENERATOR LEVEL control maximum clockwise. Allow the tubes to come up to operating temperature.
( ) Set the GENERATOR SELECTOR switch to the AC INT position.
( ) Set the DETECTOR switch to the AC INT position.
( ) Set the FUNCTION switch to the L-DQ position.
( ) Set the RANGE switch to the 1h position on the L scale.
( ) Connect a frequency counter to the top UUT (unit-under-test) terminals.
( ) Adjust the trimmer capacitor control G [chassis] (see Figure 5) until the counter reads 1000 Hz.
( ) Disconnect the frequency counter from the UUT terminals.
Front Panel Initial Conditions
( ) Set the FUNCTION switch to the R position.
( ) Set the GENERATOR switch to DC INT position.
( ) Set the DETECTION switch to DC SHUNT position.
( ) Set the RANGE switch to the 100 ohms position on the R scale.
( ) Set the outer ring of CRL control [large set of dials, right side of panel] to the center of the 5 position on the dial at the pointer position.
( ) Set the GENERATOR LEVEL control to minimum counter-clockwise, not to the AC OFF position.
* * * * * * *
CRL CALIBRATION
( ) Connect a 550-ohm 1% resistor to the UUT terminals.
( ) Set the GENERATOR LEVEL control maximum clockwise.
· The DC METER position is spring loaded and will automatically return to the DC SHUNT position.
( )* Adjust the inner CRL control only until the meter does not change when the DETECTOR switch is rotated to the DC METER position.
( )* Loosen the inner CRL control knob only and rotate it till the two dials read 5.5 at the pointer position.
( )* Tighten the inner CRL control knob.
( ) Repeat the above thee steps* until the meter does not move when the DETECTOR switch is rotated to the DC METER position.
( ) Set the GENERATOR LEVEL control counter-clockwise to the AC OFF position.
( ) Disconnect the 550-ohms 1% resistor from the UUT terminals.
* * * * * * *
DQ CALIBRATION
· If greater accuracy of the DQ control [large dial, left side of panel] is desired (more than that obtained by simply zero-setting the knob), perform the following steps:
( ) Set the front panel initial conditions as described in the OSCILLATOR ADJUSTMENT section.
( ) Temporarily disconnect the end of the wire connecting lug 2 on the rear section of the DQ control to lug 6 on the rear deck of the FUNCTION switch FS (see Figure 10).
( ) Temporarily connect a jumper wire between lug 1 of the rear section of the DQ control to lug 1 [rear UTT terminal (black)] of the terminal strip TS on top of the bridge (see Pictorial 2).
( ) Temporarily connect a jumper wire between lug 2 of the rear section of the DQ control to lug 2 [front UTT terminal (red)] of the terminal strip TS on top of the bridge (see Pictorial 2).
· These temporary connections can be wired through a set of chassis mounted toggle switches enabling ease of future DQ adjustments.
( ) Set the GENERATOR LEVEL control maximum clockwise. Allow the tubes to come up to operating temperature.
( ) Set the CRL controls to read 160 ohms.
( ) Adjust the DQ/Q control until the meter reads 0 (near the 0.1 mark on the D-Q knob scale). This is a CRITICAL adjustment.
( ) Verify the adjustment by turning the DETECTOR switch to the DC METER position. No meter pointer movement should be noticed at exact zero.
( ) When the bridge is balanced, note the position of the 0.1 mark on the DQ dial with respect to the pointer.
( )* Correct the DQ knob dial setting to read exactly 0.1.
( )* Rebalance the bridge and check the dial setting.
( ) Repeat the above two steps* until no change is recorded.
( ) Set the RANGE switch to 1000 ohms as a double check. The bridge should balance at 1.0 with the DQ dial.
( ) Set the GENERAL LEVEL control to AC OFF position.
( ) When the DQ calibration procedure is complete, remove the temporary jumpers and reconnect the proper leads to the DQ control and the FUNCTION switch.
( ) Calibration procedure is complete.