Heathkit Model IB-1B
Impedance Bridge
Calibration Procedure
· The meter pointer should be centered on the meter scale. If the pointer is not over this mark, adjust the zero-adjust screw at the bottom of the meter until the pointer is properly aligned.
· Install a fresh 6-volt battery into the unit.
Ancillary Equipment Needed
1000-ohm 1% resistor
· Mount the 1% resistor on a dual banana plug to use as a calibration standard.
* * * * * * *
Front Panel Initial Conditions
( ) Set the SELECTOR switch [top left] to the R position.
( ) Set the MULTIPLIER switch [top right] to the 1K ohms position.
( ) Set the DETECTOR switch to the shunted galv position.
( ) Set the GENERATOR switch to the DC position.
( ) Connect a 1000-ohm 1% resistor to the “R” (resistor) terminals [top left terminals].
( ) Adjust the CRL control until the galvanometer nulls or at the zero position
( ) Set the DETECTOR switch to the galv position.
* * * * * * *
RESISTANCE SCALE CALIBRATION
( ) Loosen the CRL knob carefully, checking to see that the null has not been disturbed.
( ) Set the CRL pointer indicator exactly on 1.
( ) Retighten the CRL knob carefully, checking to see that the null has not been disturbed.
( ) Set the DETECTOR switch to shunted galv position.
* * * * * * *
”D” AND “Q” SCALES CALIBRATION
· The “D” and “Q” scales will be reasonable accurate by placing the pointers on zero at the maximum counter-clockwise rotation. However, if maximum accuracy is desired, these controls may be set with the resistance section of the bridge. Otherwise, proceed to the FINAL ADJUSTMENTS section.
( ) Set the switches and controls as described in the Front Panel Initial Conditions section.
( ) Connect the “R” terminals to the “L” terminals [top right terminals] using heavy gauge short leads and leaving the 1000-ohm 1% resistor across the “R” terminals.
( ) Set the CRL pointer indicator exactly on 1.
( ) Loosen the D knob carefully, checking to see that the null has not been disturbed.
( ) Set the D pointer indicator exactly on 5.
( ) Retighten the D knob carefully, checking to see that the null has not been disturbed.
( ) Loosen the D – Q knob carefully, checking to see that the null has not been disturbed.
( ) Set the D – Q pointer indicator exactly on 5.
( ) Retighten the D - Q knob carefully, checking to see that the null has not been disturbed.
( ) Loosen the Q knob carefully, checking to see that the null has not been disturbed.
( ) Set the Q pointer indicator exactly on 50.
( ) Retighten the Q knob carefully, checking to see that the null has not been disturbed.
( ) Disconnect the heavy gauge jumper leads between the “R” terminals and the “L” terminals.
* * * * * * *
FINAL ADJUSTMENTS
( ) Set the DETECTOR switch to shunted galv position.
( ) Disconnect the 1000-ohm 1% resistor from the “R” terminals.
( ) Calibration procedure is complete.