Heathkit Model IM-4100
Frequency Counter
Calibration Procedure
Ancillary Equipment Needed
AM Radio
General Coverage Communications Receiver
Frequency Counter
RF Generator
Laboratory Standard Frequency Source
* * * * * * *
WITHOUT INSTRUMENTS
Refer to Figures 4 & 6
Illustration booklet page 4
( ) Set the POWER switch to the ON position
(Allow the unit to warm up for 30 minutes)
( ) Set the POWER/TIME BASE switch to the kHz position.
( ) Set the MODE switch to the TOL position.
“Option 1”
Using a general coverage communications receiver
( ) Set the receiver MODE switch to the AM position.
( ) Tune the receiver to the highest WWV station that can be received.
(Frequencies of: 25, 20, 15, 10, or 5 MHz)
“Option 2”
Using an AM Radio
( ) Tune the radio to a station of known frequency and of medium volume.
* * * * * * *
( ) Connect (temporarily) an insulated, unshielded wire to the receiver’s antenna terminal.
(Leave the outside antenna connected to the receiver)
( ) Lay the temporary wire over the oscillator section of the counter.
(Refer to Figure 4 in the Illustration booklet)
( ) A tone should be heard.
(If a tone is not heard, remove the case, and place the wire near IC26 - 7400)
[Left rear on the circuit board, see PICTORIAL 2-6]
( ) Adjust the oscillator trimmer through OSC ADJ hole on rear panel using a non-conducting alignment tool until a zero bear is heard from the receiver.
( ) Calibration procedure is complete.
* * * * * * *
WITH CALIBRATED INSTRUMENTS
Refer to Figure 5
Illustration booklet page 4
( ) Set the POWER switch to the ON position
(Allow the unit to warm up for 30 minutes)
( ) Set the POWER/TIME BASE switch to the kHz position.
Using a Frequency Counter and an RF Signal Generator
( ) Connect the RF Generator, reference Frequency Counter, and the Frequency Counter being calibrated as shown in Figure 5.
( ) Set the RF Generator output frequency between 1 MHz and 30 MHz.
( ) Set the RF Generator output voltage between 250 mV and 500 mV.
( ) Adjust the oscillator trimmer through OSC ADJ hole on rear panel using a non-conducting alignment tool until the IM-4100 counter reading matches the reference counter reading.
( ) Calibration procedure is complete.
Using a Known Laboratory Standard Frequency
( ) Known laboratory standard frequency should be between 1 MHz and 30 MHz
( ) Known laboratory standard frequency output voltage should be between 250 mV and 500 mV.
( ) Connect the known laboratory standard frequency to the IM-4100.
( ) Set the RF Generator output frequency between 1 MHz and 30 MHz
( ) Adjust the oscillator trimmer through OSC ADJ hole on rear panel using a non-conducting alignment tool until the IM-4100 counter reading matches the known laboratory standard frequency.
( ) Calibration procedure is complete.